Standard Operating Procedure (SOP) for Operational Check of Friability Test Apparatus

Standard Operating Procedure (SOP) for Calibration of Conductivity/TDS Meter

1.0 Purpose:

The purpose of this SOP is to outline the step-by-step procedure for conducting an operational check on the Friability Test Apparatus to ensure its proper functionality and accuracy in pharmaceutical testing.

2.0 Scope:

This SOP is applicable to all personnel involved in the operation and maintenance of the Friability Test Apparatus in the pharmaceutical laboratory.

3.0 Responsibilities:

3.1 Lab Technician:

Responsibilities of Lab Technician:

3.1 Pre-Operational Preparations:

  • 3.1.1 Ensure personal protective equipment (PPE) is worn, including lab coat and gloves.
  • 3.1.2 Verify that the Friability Test Apparatus is clean and free from any residues.
  • 3.1.3 Check and confirm the calibration status of the apparatus.

3.2 Set-Up:

  • 3.2.1 Properly place the friability drum inside the apparatus and secure it in place.
  • 3.2.2 Verify that the rotational speed control is set to the standard operating speed (usually 25 rpm).
  • 3.2.3 Ensure that the drum is adequately lubricated if required by the manufacturer.

3.3 Calibration Check:

  • 3.3.1 Load a known number of test tablets (placebo tablets or as per standard requirements) into the drum.
  • 3.3.2 Set the testing time as per the standard procedure (usually 4 minutes).
  • 3.3.3 Start the apparatus and monitor for any irregularities during the test.

3.4 Visual Inspection:

  • 3.4.1 Observe the tablets during the test to ensure there is no damage or breakage beyond the acceptable limit.
  • 3.4.2 Vigilantly monitor the drum for smooth rotation without abnormal noise or vibration.

3.5 Post-Operational Checks:

  • 3.5.1 Stop the apparatus after the completion of the test duration.
  • 3.5.2 Carefully remove the tablets from the drum.
  • 3.5.3 Inspect the tablets for any signs of excessive friability.

3.6 Documentation:

  • 3.6.1 Record the results of the operational check accurately.
  • 3.6.2 Document any observations or deviations from the standard procedure.
  • 3.6.3 Report any issues or deviations to the appropriate personnel promptly.

4.0 Review and Approval:

  • The Lab Technician is responsible for presenting the recorded results to the Quality Assurance (QA) Officer for review and approval.

3.2 Quality Assurance (QA) Officer:

Responsibilities of QA Officer:

3.2.1 Pre-Operational Preparations:

  • 3.2.1.1 Confirm that the Lab Technician has adhered to the required personal protective equipment (PPE) guidelines.
  • 3.2.1.2 Verify the cleanliness of the Friability Test Apparatus and the status of its calibration.

3.2.2 Set-Up:

  • 3.2.2.1 Ensure the Lab Technician has correctly placed the friability drum inside the apparatus and secured it in place.
  • 3.2.2.2 Verify that the rotational speed control is set to the standard operating speed.
  • 3.2.2.3 Confirm proper lubrication of the drum if required.

3.2.3 Calibration Check:

  • 3.2.3.1 Review the documented calibration check, ensuring compliance with the established standards.
  • 3.2.3.2 Check for any anomalies or deviations during the calibration check.

3.2.4 Visual Inspection:

  • 3.2.4.1 Evaluate the Lab Technician’s observations during the visual inspection of tablets for damage or breakage.
  • 3.2.4.2 Confirm the Lab Technician’s assessment of the drum’s rotation for smooth operation.

3.2.5 Post-Operational Checks:

  • 3.2.5.1 Scrutinize the Lab Technician’s documentation of the post-operational checks, including tablet inspection and any identified issues.
  • 3.2.5.2 Determine the severity of any deviations and assess the need for corrective actions.

3.2.6 Documentation Review:

  • 3.2.6.1 Thoroughly review all recorded results and documentation for completeness and accuracy.
  • 3.2.6.2 Cross-reference the findings with established quality standards and procedural requirements.

3.2.7 Approval Process:

  • 3.2.7.1 Provide final approval for the operational check if all parameters meet the specified criteria.
  • 3.2.7.2 If deviations are identified, collaborate with relevant personnel to initiate corrective actions.

4.0 Record Keeping:

  • Ensure all records related to the operational check are appropriately filed and maintained.

4.0 Equipment and Materials:

  • Friability Test Apparatus
  • Test tablets (placebo tablets or as per standard requirements)

5.0 Procedure:

5.1 Pre-Operational Preparations:

5.1.1 Personal Protective Equipment (PPE) Verification:

  • 5.1.1.1 Prior to initiating any activity, ensure that the Lab Technician verifies the use of appropriate PPE, including lab coat and gloves.
  • 5.1.1.2 Confirm that PPE is in good condition and free from any damage.

5.1.2 Apparatus Cleanliness:

  • 5.1.2.1 Inspect the Friability Test Apparatus for cleanliness and absence of residues from previous tests.
  • 5.1.2.2 If any residues are detected, clean the apparatus following the approved cleaning procedure.

5.3 Calibration Status Check:

  • 5.3.3.1 Verify the calibration status of the Friability Test Apparatus.
  • 5.3.3.2 Ensure that the calibration is up-to-date as per the established schedule.

5.4 Readiness Confirmation:

  • 5.4.4.1 Confirm that all necessary materials and tools for the operational check are available.
  • 5.4.4.2 Check the availability of the required number of placebo tablets or test tablets for the calibration check.

5.5 Safety Checks:

  • 5.5.5.1 Ensure that emergency equipment, such as fire extinguishers and first aid kits, is readily accessible.
  • 5.5.5.2 Confirm the availability and functionality of emergency exits.

5.6 Documentation Review:

  • 5.6.6.1 Check the Operational Check Log to ensure that the last operational check details and any relevant notes are reviewed.
  • 5.6.6.2 Verify that any outstanding issues from the previous operational checks are appropriately addressed.

5.2 Set-Up:

Set-Up Procedure:

5.2.1 Placement of Friability Drum:

  • Open the apparatus and place the friability drum inside.
  • Ensure the drum is securely positioned and locked in place according to the manufacturer’s guidelines.

5.2.2 Rotational Speed Control:

  • Verify that the rotational speed control is set to the standard operating speed (typically 25 rpm).
  • If adjustments are necessary, follow the equipment manual guidelines for setting the correct speed.

5.2.3 Lubrication Check:

  • Confirm whether the friability drum requires lubrication, following the manufacturer’s recommendations.
  • If lubrication is needed, apply the appropriate lubricant as instructed.

5.2.4 Power On:

  • Switch on the power to the Friability Test Apparatus.
  • Ensure that the apparatus powers up without any abnormal sounds or vibrations.

5.2.5 System Check:

  • Run a brief system check to verify that all components are functioning correctly.
  • Check for any error messages or abnormalities on the display panel.

5.2.6 Ready Status Confirmation:

  • Confirm that the apparatus is in a ready state for the operational check.
  • Ensure that the set-up process is completed in accordance with the equipment specifications.

5.2.7 Calibration Verification:

  • Double-check that the apparatus is calibrated, referring to the calibration certificate or the calibration log.
  • If the calibration is not within the acceptable range, address the issue before proceeding.

5.3 Calibration Check:

Calibration Check Procedure:

5.3.1 Load Test Tablets:

  • Place a known number of test tablets (placebo tablets or as per standard requirements) into the friability drum.
  • Confirm that the tablet load adheres to the specified quantity for the calibration check.

5.3.2 Set Testing Time:

  • Set the testing time on the Friability Test Apparatus as per the standard procedure (usually 4 minutes).
  • Ensure that the testing time is accurately inputted and displayed.

5.3.3 Initiate Apparatus:

  • Start the Friability Test Apparatus to initiate the calibration check.
  • Observe the apparatus during operation to ensure proper functioning.

5.3.4 Monitor Tablet Condition:

  • Continuously monitor the test tablets during the calibration check for signs of damage or breakage.
  • Note any deviations from the expected tablet condition.

5.3.5 Stop Apparatus:

  • Stop the apparatus after the completion of the specified testing time.
  • Ensure that the apparatus comes to a complete stop before proceeding.

5.3.6 Tablet Inspection:

  • Carefully remove the test tablets from the drum.
  • Inspect the tablets for any signs of excessive friability, such as chipping or breakage.

5.3.7 Record Results:

  • Record the results of the calibration check in the Operational Check Log.
  • Include details such as tablet condition, testing time, and any deviations observed.

5.3.8 Report Deviations:

  • If any deviations are identified during the calibration check, report them to the designated personnel immediately.
  • Document any corrective actions taken to address the deviations.

5.4 Visual Inspection:

Visual Inspection Procedure:

5.4.1 Tablet Observation:

  • Initiate the operational check of the Friability Test Apparatus as per the established procedure.
  • Continuously observe the test tablets inside the drum during the testing period.

5.4.2 Damage Assessment:

  • Evaluate each tablet for signs of damage, including chipping, breakage, or excessive wear.
  • Record any observed damage and categorize it based on severity.

5.4.3 Drum Rotation Monitoring:

  • Monitor the rotation of the friability drum for smooth and consistent movement.
  • Check for any abnormal noise or vibration that may indicate issues with the apparatus.

5.4.4 Stop Apparatus:

  • Stop the Friability Test Apparatus after the completion of the testing period.
  • Ensure that the apparatus comes to a complete stop before proceeding.

5.4.5 Tablet Removal:

  • Carefully remove the test tablets from the drum for closer inspection.
  • Handle the tablets with care to avoid additional damage during removal.

5.4.6 Detailed Inspection:

  • Examine each tablet individually for any finer details not visible during the operational check.
  • Note any additional observations or deviations discovered during the detailed inspection.

5.4.7 Record Results:

  • Record the results of the visual inspection in the Operational Check Log.
  • Include details such as observed damage, drum rotation assessment, and any deviations noted.

5.4.8 Report Deviations:

  • If any significant deviations are identified during the visual inspection, report them to the designated personnel promptly.
  • Document any corrective actions taken to address the observed deviations.

5.5 Post-Operational Checks:

Post-Operational Checks Procedure:

5.5.1 Stop Apparatus:

  • Stop the Friability Test Apparatus immediately after the completion of the specified testing time.
  • Ensure that the apparatus comes to a complete stop before proceeding with post-operational checks.

5.5.2 Tablet Removal:

  • Carefully remove the test tablets from the friability drum.
  • Handle the tablets with care to prevent additional damage during removal.

5.5.3 Visual Inspection:

  • Conduct a visual inspection of each tablet for signs of damage, chipping, or excessive wear.
  • Record any observed damage and categorize it based on severity.

5.5.4 Drum Inspection:

  • Inspect the friability drum for any residual tablet fragments or unusual wear.
  • Clean the drum if necessary, following the approved cleaning procedure.

5.5.5 Functional Checks:

  • Verify that the Friability Test Apparatus is functioning correctly, with no abnormal noise or vibrations.
  • Confirm that all components are in proper working order.

5.5.6 Review Documentation:

  • Review the documentation in the Operational Check Log for completeness and accuracy.
  • Cross-reference the results of post-operational checks with visual inspection findings.

5.5.7 Record Results:

  • Record the results of post-operational checks in the Operational Check Log.
  • Include details such as observed tablet conditions, drum inspection results, and any deviations noted.

5.5.8 Report Deviations:

  • If any significant deviations are identified during post-operational checks, report them to designated personnel promptly.
  • Document any corrective actions taken to address the observed deviations.

5.6 Documentation:

Documentation Procedure:

5.6.1 Operational Check Log:

  • Create a dedicated Operational Check Log for the Friability Test Apparatus.
  • Include fields for date, technician’s name, apparatus identification, and calibration details.

5.6.2 Pre-Operational Documentation:

  • Record the verification of personal protective equipment (PPE) usage by the Lab Technician.
  • Document the cleanliness check of the apparatus and the confirmation of its calibration status.

5.6.3 Set-Up Documentation:

  • Log details of the placement and securing of the friability drum.
  • Document the verification of rotational speed control and lubrication status.

5.6.4 Calibration Check Documentation:

  • Record the specifics of the calibration check, including the number of test tablets, testing time, and any deviations observed.
  • Document the QA Officer’s review and approval of the calibration check.

5.6.5 Visual Inspection Documentation:

  • Record observations during the visual inspection, including tablet condition and drum rotation assessment.
  • Document any issues or deviations identified during the visual inspection.

5.6.6 Post-Operational Checks Documentation:

  • Document results of post-operational checks, including tablet conditions, drum inspection, and functional checks.
  • Record any deviations noted and any corrective actions taken.

5.6.7 Record Review and Approval:

  • The Lab Technician must review recorded results for completeness and accuracy.
  • The QA Officer reviews and approves the documented results in the Operational Check Log.

5.6.8 Corrections and Amendments:

  • If errors are identified in the documentation, make corrections using the approved correction procedure.
  • Clearly indicate any amendments made and the reason for the correction.

5.6.9 Record Retention:

  • Establish a record retention system according to regulatory requirements.
  • Clearly label and file records for easy retrieval during audits.

5.7 Training and Communication:

  • Train personnel on the proper documentation procedures outlined in this SOP.
  • Communicate any updates or changes to the documentation process promptly.

6.0 Review and Approval:

  • The recorded results and any noted deviations must be reviewed and approved by the QA Officer.

7.0 Records:

Record Creation:

7.1 Operational Check Log:

  • Create a dedicated Operational Check Log for the Friability Test Apparatus.
  • Include fields for date, technician’s name, apparatus identification, and any relevant calibration details.

7.2 Pre-Operational Preparations:

  • Document the Lab Technician’s verification of personal protective equipment (PPE) usage.
  • Record the confirmation of apparatus cleanliness and calibration status.

7.3 Set-Up:

  • Log details of the correct placement and securing of the friability drum.
  • Include information on the verification of rotational speed and lubrication.

7.4 Calibration Check:

  • Record the specifics of the calibration check, such as the number of test tablets, testing time, and any deviations observed.
  • Highlight the QA Officer’s review and approval of the calibration check.

7.5 Visual Inspection:

  • Document the Lab Technician’s observations during visual inspections of tablets and drum rotation.
  • Note any issues identified during the visual inspection.

7.6 Post-Operational Checks:

  • Record the results of post-operational checks, including tablet inspections and any corrective actions taken.
  • Include details of the QA Officer’s review and approval of the post-operational checks.

7.7 Documentation Review:

  • Regularly review records for completeness, accuracy, and adherence to established standards.
  • Document any discrepancies or corrective actions taken during the review process.

7.8 Record Retention:

  • Establish a systematic filing system for easy retrieval and long-term storage of operational check records.
  • Define a retention period for records in accordance with regulatory requirements.

7.9 Accessibility:

  • Ensure that all relevant personnel have access to the records when necessary.
  • Implement security measures to protect the confidentiality and integrity of the records.

8.0 Training and Communication:

  • Train personnel on the proper documentation procedures outlined in this SOP.
  • Communicate any updates or changes to the records management process promptly.

8.0 References:

  • Include any relevant reference documents or standards applicable to the friability testing.

9.0 Revision History:

  • Maintain a record of revisions made to this SOP.

10.0 Attachments:

Attach any supporting documents, calibration certificates, or test procedures relevant to the Friability Test Apparatus.

For more articles, Kindly Click here

For pharmaceutical jobs, follow us on LinkedIn

For more information kindly follow us on pharmaguidelines.co.uk

One Reply to “Standard Operating Procedure (SOP) for Operational Check of Friability Test Apparatus”

  1. The breadth of knowledge compiled on this website is astounding. Every article is a well-crafted masterpiece brimming with insights. I’m grateful to have discovered such a rich educational resource. You’ve gained a lifelong fan!

Leave a Reply

Your email address will not be published. Required fields are marked *