Standard Operating Procedure (SOP) for Calibration Procedure of U.V. Cabinet

Standard Operating Procedure (SOP) for Calibration of Conductivity/TDS Meter

1.0 Purpose

The purpose of this Standard Operating Procedure (SOP) is to establish a systematic and standardized approach for the calibration of the U.V. (Ultraviolet) Cabinet to ensure accurate and reliable performance in accordance with regulatory requirements.

2.0 Scope

This SOP applies to all personnel involved in the calibration of U.V. Cabinets within the organization. It covers the procedures for calibration, documentation, and the frequency of calibration activities.

3.0 Responsibilities

3.1 Laboratory Manager:

Responsibilities of the Laboratory Manager

3.1 Authorization and Oversight

  • The Laboratory Manager is responsible for authorizing all U.V. Cabinet calibration activities within the laboratory.
  • Oversee the entire calibration process to ensure compliance with this SOP and regulatory requirements.

3.2 Resource Allocation

  • Allocate necessary resources, including personnel and equipment, to facilitate smooth and effective U.V. Cabinet calibration.

3.3 Calibration Schedule Management

  • Develop and maintain a schedule for U.V. Cabinet calibrations, ensuring that they are conducted at the required frequency as per manufacturer recommendations and regulatory standards.

3.2 Calibration Technician:

Responsibilities of the Calibration Technician

3.1 Pre-Calibration Checks

  • Conduct thorough pre-calibration checks, ensuring the U.V. Cabinet is turned off and inspecting for any physical damage or irregularities.
  • Verify the cleanliness of the cabinet and the optimal condition of U.V. lamps.

3.2 Calibration Setup

4.0 Equipment and Materials

4.1 U.V. Cabinet

  • Ensure that the U.V. Cabinet is clean and free from any contaminants before starting the calibration process.
  • Ensure that the U.V. lamps are in proper working condition.

4.2 Calibration Standards

  • Use calibrated reference standards traceable to national or international standards.

4.3 Measurement Instruments

  • Calibrated measurement instruments such as UV radiometers.

5.0 Calibration Procedure

5.1 Pre-Calibration Checks

  • Verify that the U.V. Cabinet is turned off.
  • Inspect the cabinet for any physical damage or irregularities.
  • Check the cleanliness of the cabinet and the condition of U.V. lamps.

5.2 Calibration Setup

  • Place the calibrated reference standard inside the U.V. Cabinet.
  • Ensure that the reference standard is positioned in the center for uniform exposure.
  • Turn on the U.V. Cabinet and allow it to stabilize for at least 15 minutes.

5.3 Measurement

  • Use the calibrated UV radiometer to measure the U.V. intensity at various points within the cabinet.
  • Record the measurements systematically, ensuring coverage of the entire working area.

5.4 Adjustments

  • If the measurements deviate from the acceptable range, make necessary adjustments to the U.V. Cabinet settings following manufacturer guidelines.

5.5 Post-Calibration Checks

  • Re-measure the U.V. intensity at the same points to ensure that adjustments were effective.
  • Document the post-calibration measurements.

6.0 Documentation

6.1 Calibration Log

The purpose of this Standard Operating Procedure (SOP) is to establish a standardized process for maintaining a comprehensive Calibration Log during the calibration procedure of U.V. Cabinets. This ensures accurate recording of calibration activities and adherence to regulatory requirements.

7.0 Frequency of Calibration

The purpose of this Standard Operating Procedure (SOP) is to define the frequency at which U.V. Cabinets will undergo calibration, ensuring their continued accuracy, reliability, and compliance with regulatory standards.

any changes made to this SOP, including the date of revision and the person authorizing the revision.

8.0 Training

Ensure that personnel involved in the U.V. Cabinet calibration process are adequately trained and competent to perform the calibration activities outlined in this SOP.

9.0 References

List any relevant standards, manufacturer manuals, or regulatory requirements applicable to U.V. Cabinet calibration.

10.0 Records Retention

Retain calibration records for a minimum of [Specify duration based on regulatory requirements].

11.0 Revision History

Maintain a revision history section to track any changes made to this SOP, including the date of revision and the person authorizing the revision.

For more articles, Kindly Click here.

For pharmaceutical jobs, follow us on LinkedIn 

For Editable SOPs in word format contact us on info@pharmaceuticalcarrier.com 

For more information kindly follow us on www.pharmaguidelines.co.uk

Leave a Reply

Your email address will not be published. Required fields are marked *